  • Eight-channel-LED(1).jpg
  • 8 kou.png

Eight-channel LED Light Source

ORI-KJ-07-FE01 is a high-power eight-channel LED system for applications such as microscope illumination, multi-color fluorophore excitation. The device consists of eight high-power LED light sources that are collinear and output by a liquid-core optical waveguide, which can realize the characteristics of flexible wavelength, convenient use, and stable output power. In addition, this eight-channel LED light source can realize continuous or millisecond-level pulse output through the corresponding driver, and the light intensity can be continuously adjusted. It can also be remotely control through computer. We support the secondary development of customers.


Product Classification:

LED fluorescence light source


key word:

Eight-channel

LED Light Source

Fluorescent Imaging

LED microscope lighting

fluorescence microscopy


Product Details


ORI-KJ-07-FE01 is a high-power eight-channel LED system for applications such as microscope illumination, multi-color fluorophore excitation. The device consists of eight high-power LED light sources that are collinear and output by a liquid-core optical waveguide, which can realize the characteristics of flexible wavelength, convenient use, and stable output power. In addition, this eight-channel LED light source can realize continuous or millisecond-level pulse output through the corresponding driver, and the light intensity can be continuously adjusted. It can also be remotely control through computer. We support the secondary development of customers.

Table1. Parameters Table

Category

Eight-Channel LED System

Wavelength

370nm420nm470nm525nm595nm630nm660nm730nm

Optical Power

Details as bellowed table 2

Output Type

Liquid core optical waveguide

Lifetime

>10000h

Operating Temperature

0~40

Storage Temperature

-40~70

Dimension

240×181×100mm

Weight

3580g

 

Operating Environment

The LED will gradually heat up when using, which will lead to the attenuation of optical power and even shorten the life of the LED. We have excellent heat dissipation structure design, which can prolong the lifetime of LED light source and reduce power loss. Please work in a ventilated environment to avoid overheating, to avoid the decrease of optical power and shorten the lifetime of LED.

Table 2. LED Typical Optical Characteristics

Peak Wavelengthnm

Rangenm

Radiated Optical PowermW

Spectral FWHMnm

370

365-370

300

16

420

415-425

350

20

465

465-470

390

32

525

520-530

80

30

595

590-600

70

16

635

630-640

190

20

660

655-665

460

16

730

725-735

250

34

 

1.The wavelength error is ±5.0nm, and the radiated optical power error is ±5%. The above parameters are for reference only, and the actual factory test data shall prevail.

2. All Specifications subject to change.

Power Supply

We provide the corresponding light source driver, as shown in the figure below. If you are equipped with a constant current drive power supply, please contact us for the pin diagram and strictly follow it when connecting. At the same time, please ensure that the operating current does not exceed the maximum allowable value, provide sufficient forward voltage, and connect the pins correctly.

Fiber

Multimode fiber bundles

Bundles of 7 multimode fibers

SMA905 connectors on both ends

Stainless steel tubing

Liquid light guides

Transmits light with great intensity

Flexible and unbreakable

Large aperture about twice the size of silica fiber light guides

Maintenance & Service

To avoid damage, do not expose them to sprays, liquids or solvents, and do not open the outer covering. In case of failure, please contact service@oricszlab.com, we will provide you with after-sales service and technical support as soon as possible.

Warning and safety

This LED is not suitable for home interior lighting, must not be operated in explosive atmospheres, and can only be used with shielded connecting cables.

UV Warning

The LED emits strong light when working. Part of the LED emission spectrum contains UV components. Do not look directly at the LED light during operation, even a short exposure time may be harmful to the eyes. If direct viewing of the beam is required, protective goggles must be worn to avoid eye injury.

Pre

Next

<Pre

Next>

GET IN TOUCH

Submit Message